Comparative evaluation of six rapid serological tests for HIV-1 antibody.

  • 1 February 1993
    • journal article
    • research article
    • Vol. 6 (2), 115-9
Abstract
Six commercial rapid test kits for HIV-1 antibody were evaluated. Four laboratory technologists tested sera from four groups of U.S. military personnel or dependents: 202 subjects positive for HIV-1 antibody by Western blot, 200 seronegative voluntary blood donors, 199 seronegative obstetrics/gynecology patients, and 99 subjects with sera reactive by ELISA but negative by recombinant protein EIA and indeterminate by Western blot. The three tests using solid-phase immunoassay technology demonstrated the highest mean sensitivity (> 99%) and specificity (> 91%) for all groups tested, including sera indeterminate by Western blot. Two dot-immunoblot assays were less specific, possibly due to indistinct reaction end points, and a latex agglutination assay was also less specific because of difficulty distinguishing reactive results from the granular background. In an "ease-of-use" assessment, solid-phase capture immunoassays required less time and equipment and were easier to interpret than other testing methods. Solid-phase capture immunoassays for HIV-1 antibody may be suitable for use in emergency situations and in developing countries because they are highly sensitive and specific and are rapidly performed with minimal laboratory equipment.