Statistical validation of reagent lot change in the clinical chemistry laboratory can confer insights on good clinical laboratory practice
Open Access
- 4 February 2014
- journal article
- research article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 51 (6), 688-694
- https://doi.org/10.1177/0004563214520749
Abstract
Verification of new lot reagent’s suitability is necessary to ensure that results for patients’ samples are consistent before and after reagent lot changes. A typical procedure is to measure results of some patients’ samples along with quality control (QC) materials. In this study, the results of patients’ samples and QC materials in reagent lot changes were analysed. In addition, the opinion regarding QC target range adjustment along with reagent lot changes was proposed. Patients’ sample and QC material results of 360 reagent lot change events involving 61 analytes and eight instrument platforms were analysed. The between-lot differences for the patients’ samples (ΔP) and the QC materials (ΔQC) were tested by Mann–Whitney U tests. The size of the between-lot differences in the QC data was calculated as multiples of standard deviation (SD). The ΔP and ΔQC values only differed significantly in 7.8% of the reagent lot change events. This frequency was not affected by the assay principle or the QC material source. One SD was proposed for the cutoff for maintaining pre-existing target range after reagent lot change. While non-commutable QC material results were infrequent in the present study, our data confirmed that QC materials have limited usefulness when assessing new reagent lots. Also a 1 SD standard for establishing a new QC target range after reagent lot change event was proposed.Keywords
This publication has 13 references indexed in Scilit:
- Commutability Limitations Influence Quality Control Results with Different Reagent LotsClinical Chemistry, 2011
- Quality ControlPublished by Elsevier BV ,2011
- State of the Art in Trueness and Interlaboratory Harmonization for 10 Analytes in General Clinical ChemistryArchives of Pathology & Laboratory Medicine, 2008
- Comparison of Fresh Frozen Serum to Traditional Proficiency Testing Material in a College of American Pathologists Survey for Ferritin, Folate, and Vitamin B12Archives of Pathology & Laboratory Medicine, 2005
- Creatinine Measurement: State of the Art in Accuracy and Interlaboratory HarmonizationArchives of Pathology & Laboratory Medicine, 2005
- Comparison of Pooled Fresh Frozen Serum to Proficiency Testing Material in College of American Pathologists Surveys: Cortisol and Immunoglobulin EArchives of Pathology & Laboratory Medicine, 2005
- Specimen materials, target values and commutability for external quality assessment (proficiency testing) schemesClinica Chimica Acta; International Journal of Clinical Chemistry, 2003
- Trueness verification in European external quality assessment schemes: time to care about the quality of the samples.Scandinavian Journal of Clinical and Laboratory Investigation, 2003
- Impact of reference materials on accuracy in clinical chemistryClinical Biochemistry, 1998
- Quality control in clinical chemistry: characterization of reference materialsTalanta, 1984