New torch design with an in-built chamber for liquid sample analysis by ICP-AES
- 7 March 2002
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 17 (4), 345-351
- https://doi.org/10.1039/b111416f
Abstract
The analytical performance of the torch integrated sample introduction system (TISIS) for the analysis of very low sample volumes by ICP-AES is described. The TISIS consists of a commercially available torch with a low inner volume cavity (i.e., 5–9 cm3) placed at its base, which is attached to the shortened plasma injector tube through a special PTFE adapter. The nebulizer is vertically fitted to the torch base by means of a home-made Teflon adapter that has an exit for the drain solution. The aerosol is thus generated at the base of the TISIS cavity. The nebulization conditions corresponded to very low liquid flow rates (i.e., from 5 to 200 µl min−1) and conventional gas flow rates (i.e., 0.7 l min−1). Based on previously published work, these nebulization conditions favored aerosol solvent evaporation. At ambient temperature and below 40 µl min−1, almost all the solvent was evaporated and no drain was needed. Therefore, the main purpose of the TISIS cavity was to promote efficient solvent evaporation (instead of selecting the maximum drop size introduced into the plasma) and to simplify the transport path. In order to enhance solvent evaporation, the primary aerosols must be as fine as possible. In the present work, two pneumatic micronebulizers were employed: a so-called high efficiency nebulizer (HEN) and a MicroMist. For the sake of comparison, the HEN was coupled to a double-pass spray chamber and the MicroMist to a cyclonic type spray chamber. The results indicated that the TISIS exhibited promising behavior in ICP-AES, giving rise to emission intensities similar to or higher than both conventional reference systems. No further deterioration of the plasma thermal characteristics was observed when using the TISIS with respect to conventional chambers. The signal stability and limits of detection were similar for the different systems employed.Keywords
This publication has 33 references indexed in Scilit:
- Influence of the spray chamber design for vapor-based liquid sample introduction at room temperature in ICP-AESJournal of Analytical Atomic Spectrometry, 2002
- Non-spectral interference effects in inductively coupled plasma mass spectrometry using direct injection high efficiency and microconcentric nebulisationJournal of Analytical Atomic Spectrometry, 2000
- Application of double-focusing sector field ICP mass spectrometry with shielded torch using different nebulizers for ultratrace and precise isotope analysis of long-lived radionuclidesJournal of Analytical Atomic Spectrometry, 1999
- Comparison of characteristics and limits of detection of pneumatic micronebulizers and a conventional nebulizer operating at low uptake rates in ICP-AESJournal of Analytical Atomic Spectrometry, 1999
- Improving sensitivity for CE-ICP-MS using multicapillary parallel separationJournal of Analytical Atomic Spectrometry, 1999
- Dispersion of discrete sample signals within aerosol spray chambers: preliminary investigationsSpectrochimica Acta Part B: Atomic Spectroscopy, 1996
- Influence of operating conditions on the effects of acids in inductively coupled plasma atomic emission spectrometryJournal of Analytical Atomic Spectrometry, 1994
- Small spray chamber for inductively coupled plasma atomic emission spectrometry and its evaluation by a digital signal sampling techniqueJournal of Analytical Atomic Spectrometry, 1989
- Direct atomic spectrometric analysis by slurry atomisation. Part 8. Flow injection inductively coupled plasma atomic emission spectrometryJournal of Analytical Atomic Spectrometry, 1989
- Pneumatic nebulisers and spray chambers for inductively coupled plasma spectrometry. A review. Part 2. Spray chambersJournal of Analytical Atomic Spectrometry, 1988