Convex Lens-Induced Confinement for Imaging Single Molecules
- 17 June 2010
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 82 (14), 6224-6229
- https://doi.org/10.1021/ac101041s
Abstract
Fluorescence imaging is used to study the dynamics of a wide variety of single molecules in solution or attached to a surface. Two key challenges in this pursuit are (1) to image immobilized single molecules in the presence of a high level of fluorescent background and (2) to image freely diffusing single molecules for long times. Strategies that perform well by one measure often perform poorly by the other. Here, we present a simple modification to a wide-field fluorescence microscope that addresses both challenges and dramatically improves single-molecule imaging. The technique of convex lens-induced confinement (CLIC) restricts molecules to a wedge-shaped gap of nanoscale depth, formed between a plano-convex lens and a planar coverslip. The shallow depth of the imaging volume leads to 20-fold greater rejection of background fluorescence than is achieved with total internal reflection fluorescence (TIRF) imaging. Elimination of out-of-plane diffusion leads to an approximately 10 000-fold longer diffusion-limited observation time per molecule than is achieved with confocal fluorescence correlation spectroscopy. The CLIC system also provides a new means to determine molecular size. The CLIC system does not require any nanofabrication, nor any custom optics, electronics, or computer control.Keywords
This publication has 26 references indexed in Scilit:
- DNA manipulation, sorting, and mapping in nanofluidic systemsChemical Society Reviews, 2010
- Single Molecule Nanocontainers Made Porous Using a Bacterial ToxinJournal of the American Chemical Society, 2009
- Droplets for Ultrasmall-Volume AnalysisAnalytical Chemistry, 2009
- Controlling Brownian motion of single protein molecules and single fluorophores in aqueous bufferOptics Express, 2008
- Suppressing Brownian motion of individual biomolecules in solutionProceedings of the National Academy of Sciences of the United States of America, 2006
- Single-molecule Förster resonance energy transfer study of protein dynamics under denaturing conditionsProceedings of the National Academy of Sciences of the United States of America, 2005
- Total internal reflection fluorescence correlation spectroscopy (TIR-FCS) with low background and high count-rate per moleculeOptics Express, 2005
- Zero-Mode Waveguides for Single-Molecule Analysis at High ConcentrationsScience, 2003
- Immobilization in Surface-Tethered Lipid Vesicles as a New Tool for Single Biomolecule SpectroscopyThe Journal of Physical Chemistry B, 2001
- Total Internal Reflection Fluorescence Microscopy in Cell BiologyTraffic, 2001