Diffusion Tensor Imaging of the Kidney With Parallel Imaging: Initial Clinical Experience
- 1 October 2008
- journal article
- Published by Ovid Technologies (Wolters Kluwer Health) in Investigative Radiology
- Vol. 43 (10), 677-685
- https://doi.org/10.1097/rli.0b013e31817d14e6
Abstract
Objective: To evaluate the clinical feasibility of diffusion tensor imaging (DTI) of the kidney in volunteers and patients with renal diseases. Material and Methods: Ten volunteers and 22 patients (mean age, 56 ± 14.3) with renal masses and renal artery stenosis underwent breath-hold coronal fat-saturated echo-planar DTI (as provided by the manufacturer, 6 diffusion directions, diffusion weightings b = 0 and 300 s/mm2, repetition time 730 ms/echo time 72 ms; 5 slices; slice thickness, 6 mm; inplane resolution, 2.1 × 2.1 mm2; acquisition time, 26 seconds) of the kidneys at 1.5 T (MAGNETOM Avanto, Siemens Medical Solutions, Erlangen, Germany). The parallel imaging technique, generalized autocalibrating partially parallel acquisitions with an acceleration factor 2, was applied. Using the commercially available Syngo DTI task card software, regions of interests were placed in the cortex, medulla, and in renal masses if present. Fractional anisotropy (FA) and apparent diffusion coefficients (ADC) were determined, and tractography was used to visualize the renal diffusion properties. Statistical analysis was performed using the Wilcoxon signed-rank sum test and paired t tests. Results: In all volunteers, FA was significantly (P < 0.01) higher in the medulla (0.36 ± 0.03) than in the cortex (0.21 ± 0.02), whereas the ADC was significantly (P < 0.01) higher in the cortex (2.43 ± 0.19) than in the medulla (2.16 ± 0.22). Tractography typically revealed a radial preferred direction of medullary diffusion basically reflecting medullary flow. FA/ADC of simple renal cysts (n = 8) was 0.14 ± 0.05/2.86 ± 0.15. Renal cell carcinoma (n = 10) showed a wide FA range from 0.11 to 0.56. Using tractography, the structural organization of renal cell carcinoma such as pseudocapsules could be visualized. In 1 patient with unilateral high-grade renal artery stenosis, the cortical ADC of the affected kidney was lower than on the contralateral side (1.77/2.27) and the FA was increased (0.33/0.18). The FA of the medulla was increased (0.70/0.41) and the ADC decreased (1.43/1.90). Conclusions: Using parallel imaging, DTI measurements of the kidneys are feasible within a single breath-hold with good discrimination between cortex and medulla. Parallel imaging allows more slices and a superior resolution. DTI measurements of the kidney allows visualization of medullary flow, in pathology ADC and FA were altered. Further investigations will be required to evaluate the role of DTI for studying and monitoring renal ultrastructure.Keywords
This publication has 30 references indexed in Scilit:
- Magnetic Resonance Imaging in Acute and Chronic Kidney Diseases: Present StatusNephron Clinical Practice, 2006
- Diffusion-weighted MR Imaging of Kidneys in Healthy Volunteers and Patients with Parenchymal Diseases: Initial ExperienceRadiology, 2005
- Diffusion-weighted MRI in the evaluation of renal lesions: preliminary resultsThe British Journal of Radiology, 2004
- Single breath-hold diffusion-weighted imaging of the abdomenJournal of Magnetic Resonance Imaging, 2003
- Renal diffusion and BOLD MRI in experimental diabetic nephropathyJournal of Magnetic Resonance Imaging, 2002
- From the diffusion coefficient to the diffusion tensorNMR in Biomedicine, 2002
- Diffusion tensor MRI of the human kidneyJournal of Magnetic Resonance Imaging, 2001
- Anisotropic diffusion in kidney: Apparent diffusion coefficient measurements for clinical useJournal of Magnetic Resonance Imaging, 2000
- Diffusion MR imaging: clinical applications.American Journal of Roentgenology, 1992
- MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders.Radiology, 1986