Comprehensive characterization and resolution of discrepant spectrophotometric bilirubin results in patients on eltrombopag therapy
Open Access
- 1 October 2020
- journal article
- research article
- Published by Walter de Gruyter GmbH in cclm
- Vol. 58 (10), 1713-1723
- https://doi.org/10.1515/cclm-2019-0684
Abstract
Background: Eltrombopag is a thrombopoietin receptor agonist used for the treatment of thrombocytopenic conditions. It can cause pH-dependent discoloration of plasma/serum. Eltrombopag is potentially hepatotoxic. It can affect the assessment of hyperbilirubinemia because of its (i) absorbance at similar to 450 nm (bilirubin), (ii) absorbance at similar to 550 nm (diazo-bilirubin) and (iii) it can cause yellowish discoloration of the eyes at normal circulating bilirubin levels. Methods: We collected 66 samples from patients on a range of eltrombopag dosages up to 150 mg daily. Bilirubin was measured using multiple routine spectrophotometric analyzers, the Doumas reference method and high-performance liquid chromatography (HPLC). Plasma/serum eltrombopag concentrations were determined using liquid chromatography tandem mass spectrometry (LC-MS/MS). Spike-in and admixture experiments delineated the effects of eltrombopag and its metabolites. Results: Forty-nine of 52 samples from patients on >= 50 mg daily eltrombopag therapy showed significantly discrepant inter-analyzer total bilirubin results, a difference up to 64 mu mol/L (3.7 mg/dL). In one sample, total bilirubin varied from 8 to 65 mu mol/L (0.4-3.8 mg/dL) by different routine analyzers, with direct bilirubin <= 4 mu mol/L (0.2 mg/dL). There was a positive correlation between total bilirubin difference and plasma eltrombopag concentration (r= 0.679), and spike-in experiments demonstrated that Beckman AU and Doumas reference methods were susceptible to positive interference. HPLC can quantify bilirubin after separating eltrombopag, and results suggest different analyzers are affected to varying degrees by eltrombopag and its metabolites. Conclusions: Eltrombopag and its metabolites can cause positive interference to the spectrophotometric measurements of total bilirubin. Accurate measurements of total bilirubin may improve our understanding of the prevalence of hyperbilirubinemia in patients on eltrombopag therapy.This publication has 20 references indexed in Scilit:
- Eltrombopag: liver toxicity, kidney injury or assay interference?Pathology, 2016
- Eltrombopag interference in routine chemistry testingAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 2016
- Eltrombopag: Investigation of Previously Reported Serum Discoloration and Negative Interference on a Total Bilirubin Diazo MethodArchives of Pathology & Laboratory Medicine, 2016
- Liquid chromatography–tandem mass spectrometric assay for eltrombopag in 50μL of human plasma: A pharmacokinetic studyJournal of Pharmaceutical and Biomedical Analysis, 2014
- Eltrombopag and Serum of a Different HueArchives of Pathology & Laboratory Medicine, 2013
- Metabolism and Disposition of Eltrombopag, an Oral, Nonpeptide Thrombopoietin Receptor Agonist, in Healthy Human SubjectsDrug Metabolism and Disposition, 2011
- Population Pharmacokinetics of Eltrombopag in Healthy Subjects and Patients With Chronic Idiopathic Thrombocytopenic PurpuraThe Journal of Clinical Pharmacology, 2011
- Effects of food and antacids on the pharmacokinetics of eltrombopag in healthy adult subjects: Two single-dose, open-label, randomized-sequence, crossover studiesClinical Therapeutics, 2009
- Accurate and sensitive high-performance liquid chromatographic method for geometrical and structural photoisomers of bilirubin IXα using the relative molar absorptivity valuesJournal of Chromatography A, 1999
- Candidate reference method for determination of total bilirubin in serum: development and validation.Clinical Chemistry, 1985