Step-Response-Based Calibration Method for ESD Generators in the Air-Discharge Mode

Abstract
The current IEC 61000-4-2 (2008) standard does not provide calibration specifications for electrostatic discharge (ESD) generators in the air-discharge mode. This is largely due to the well-known fact of poor repeatability in air-discharge measurements. This complicates identifying an acceptable tolerance range of a reference waveform for air discharge. The variability of the discharge waveforms is caused by variations in the spark length. A novel air-discharge calibration method is proposed, which avoids sparking. The method is based on measuring the step response of an ESD generator in the air-discharge mode using a mercury-wetted relay. Possible nonlinear effects are identified during the contact-mode ESD simulator calibration at higher voltages; thus, in combination, an air-discharge calibration method is provided. This letter explains the method and shows excellent repeatability.

This publication has 14 references indexed in Scilit: