Standardization of Clinical Cholinesterase Measurements

Abstract
Previous studies showed that commonly used kits for measuring cholinesterases were not optimal for determining acetyl-cholinesterase (ACh E) activity. Clinical use of different kits and methodologies resulted in ACh E levels being reported in different units and activities that were not reproducible among laboratories. Findings such as these led to a revision in California regulations (covering ACh E measurements for pesticide worker safety) calling for clinical laboratories to standardize their findings. The laboratories were contacted and invited to participate in a split-sample study of human blood ACh E and nonspecific cholinesterase (BCh E) assays. Participating laboratories measured erythrocyte (RBC) ACh E and/or plasma BCh E from undiluted and 50% diluted blood, according to their practices. Aliquots of blood samples were shipped to University of California Davis for measurement, using an optimized semiautomated plate reader version of the method of Ellman. Nine of 25 laboratories sent samples for comparison. Two others performed their own comparisons and submitted data to the state. Best correlations were obtained with BCh E activity. Correlations ( r2) were.88 or above for four of five laboratories for BCh E, and above.9 for two of seven laboratories for ACh E. Reasons for poor correlations may include difficulties in pipetting RBCs, storage, and processing. A bovine ACh E RBC ghost “standard” was devised and tested. Activity of the preparation was maintained at-70°C for approximately 11 months. A test with an East coast laboratory resulted in a high correlation, demonstrating the reliability of the RBC ghost standard and that one laboratory can replicate the ACh E findings of another. The overall poor correlation of interlaboratory cholinesterase results points to the need to further standardize sample handling and assay methods.

This publication has 4 references indexed in Scilit: