The Use of Stepper Motor-Controlled Proportional Valve for Fio2 Calculation in the Ventilator and its Control with Fuzzy Logic
Top Cited Papers
- 5 November 2016
- journal article
- systems level-quality-improvement
- Published by Springer Science and Business Media LLC in Journal of Medical Systems
- Vol. 41 (1), 1
- https://doi.org/10.1007/s10916-016-0650-y
Abstract
This article proposes the employment of a proportional valve that can calculate the amount of oxygen in the air to be given to patient in accordance with the amount of FiO2 which is set from the control menu of the ventilation device. To actualize this, a stepper motor-controlled proportional valve was used. Two counts of valves were employed in order to control the gases with 2 bar pressure that came from both the oxygen and medical air tanks. Oxygen and medical air manometers alongside the pressure regulators were utilized to perform this task. It is a fuzzy-logic-based controller which calculates at what rate the proportional valves will be opened and closed for FiO2 calculation. Fluidity and pressure of air given by the ventilation device were tested with a FlowMeter while the oxygen level was tested using the electronic lung model. The obtained results from the study revealed that stepper motor controlled proportional valve could be safely used in ventilation devices. In this article, it was indicated that fluidity and pressure control could be carried out with just two counts of proportional valve, which could be done with many solenoid valves, so this reduces the cost of ventilator, electrical power consumed by the ventilator, and the dimension of ventilator.Keywords
This publication has 15 references indexed in Scilit:
- Design and Construction of a Microcontroller-Based Ventilator Synchronized with Pulse OximeterJournal of Medical Systems, 2016
- Integrated micro power frequency breath detectorSensors and Actuators A: Physical, 2016
- An interactive and multi-functional refreshable Braille device for the visually impairedDisplays, 2016
- Gas transfer model to design a ventilator for neonatal total liquid ventilationMedical Engineering & Physics, 2015
- Measurements of the density and viscosity of 1-hexene+1-octene mixtures at high temperatures and high pressuresThermochimica Acta, 2014
- FiO2 Delivered By a Turbine Portable Ventilator with an Oxygen Concentrator in an Austere EnvironmentThe Journal of Emergency Medicine, 2014
- Ventilation Practices in the Neonatal Intensive Care Unit: A Cross-Sectional StudyThe Journal of Pediatrics, 2010
- Design and Realization of a Microcontroller Based E-Test Strip Application DeviceInstrumentation Science & Technology, 2009
- A Novel Fuzzy Logic Inference System for Decision Support in Weaning from Mechanical VentilationJournal of Medical Systems, 2009
- The design of thermoelectric footwear heating system via fuzzy logic.Journal of Medical Systems, 2007