Immunoassay Interference by a Commonly Used Blood Collection Tube Additive, the Organosilicone Surfactant Silwet L-720
Open Access
- 1 October 2005
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 51 (10), 1874-1882
- https://doi.org/10.1373/clinchem.2005.055400
Abstract
Background: A small number of immunoassays on several different types of analyzers were recently adversely affected by tube additives in Becton Dickinson (BD) Vacutainer® SST™, SST II, and Microtainer™ blood collection tubes. We examined the effect of a commonly used tube surfactant, Silwet™ L-720, on immunoassays and the mechanism for the interference. Methods: Immunoassays were performed on serum supplemented with Silwet L-720 on the IMMULITE™ 2500 and AxSYM™ analyzers. Direct effects of the surfactant on the chemiluminescent detection step of immunoassays and on antibody immobilization on the solid phase were examined. Results: Increasing the final surfactant concentration from 0 to 400 mg/L in serum significantly increased (∼51%) the apparent total triiodothyronine (TT3) concentrations measured on the IMMULITE 2500 but not the AxSYM analyzer. Several other competitive, but not noncompetitive, assays were also significantly affected by the surfactant on the IMMULITE 2500 analyzer. The effect was independent of serum components, and the surfactant had no direct effect on chemiluminescence reactions. The capture antibody, however, was displaced from the solid phase by incubation with solutions containing surfactant under conditions similar to the IMMULITE TT3 assay. Conclusions: The Silwet L-720 surfactant, which is used to coat the inner surfaces of tubes, appears to account for previously reported immunoassay interference by BD Vacutainer SST blood collection tubes. One of the mechanisms for the interference is the desorption of antibodies from the solid phase by the surfactant. The results identify an important factor in the selection of suitable blood collection tube surfactants and provide an approach for solving similar tube-assay interference problems in the future.Keywords
This publication has 22 references indexed in Scilit:
- Effect of Blood Collection Tubes on Total Triiodothyronine and Other Laboratory AssaysClinical Chemistry, 2005
- False Positive Results for the AUSZYME Monoclonal TestObstetrics & Gynecology, 2005
- Potential Interferences from Blood Collection Tubes in Mass Spectrometric Analyses of Serum PolypeptidesClinical Chemistry, 2004
- Evaluation of AVL988/4 analyzer for measurement of ionized magnesium and ionized calciumScandinavian Journal of Clinical and Laboratory Investigation, 2001
- Demonstrations of chemiluminescenceMethods in Enzymology, 2000
- Measurement of ionized magnesium with AVL 988/4 electrolyte analyzer: Preliminary analytical and clinical resultsScandinavian Journal of Clinical and Laboratory Investigation, 1996
- More on the measurement of ionized magnesium in whole bloodScandinavian Journal of Clinical and Laboratory Investigation, 1996
- A new plastic evacuated tube with plasma separatorJournal of Clinical Laboratory Analysis, 1995
- Absorption of Therapeutic Drugs by Barrier Gels in Serum Separator Blood Collection Tubes:Volume- and Time-dependent Reduction in Total and Free Drug ConcentrationsaAmerican Journal of Clinical Pathology, 1994
- Measurement of ionized magnesium with neutral carrier based ISE's. Progress and results with the AVL 988-4 magnesium analyzerScandinavian Journal of Clinical and Laboratory Investigation, 1994