Portal dosimetry of small unflattened beams
- 26 May 2021
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 66 (11), 11NT01
- https://doi.org/10.1088/1361-6560/ab843d
Abstract
We developed and validated a dedicated small field back-projection portal dosimetry model for pretreatment and in vivo verification of stereotactic plans entailing small unflattened photon beams. For this purpose an aSi-EPID was commissioned as a small field dosimeter. Small field output factors for 6 MV FFF beams were measured using the PTW microDiamond detector and the Agility 160-leaf MLC from Elekta. The back-projection algorithm developed in our department was modified to better model the small field physics. The feasibility of small field portal dosimetry was validated via absolute point dose differences w.r.t. small static beams, and 5 hypofractionated stereotactic VMAT clinical plans measured with the OCTAVIUS 1000 SRS array dosimeter and computed with the treatment planning system Pinnacle v16.2. Dose reconstructions using the currently clinically applied back-projection model were also computed for comparison. We found that the latter yields underdosage of about -8% for square beams with cross section near 10 mm x 10 mm and about -6% for VMAT treatments with PTV volumes smaller than about 2cm(3). With the methods described in this work such errors can be reduced to less than the +/- 3.0% recommendations for clinical use. Our results indicate that aSi-EPIDs can be used as accurate small field radiation dosimeters, offering advantages over point dose detectors, the correct positioning and orientation of which is challenging for routine clinical QA.This publication has 37 references indexed in Scilit:
- A synthetic diamond diode in volumetric modulated arc therapy dosimetryMedical Physics, 2013
- In aqua vivo EPID dosimetryMedical Physics, 2011
- Current status and future perspective of flattening filter free photon beamsMedical Physics, 2011
- A contemporary review of stereotactic radiotherapy: Inherent dosimetric complexities and the potential for detrimentActa Oncologica, 2011
- 3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetryRadiotherapy and Oncology, 2010
- The influence of small field sizes, penumbra, spot size and measurement depth on perturbation factors for microionization chambersPhysics in Medicine & Biology, 2009
- The long‐term stability of amorphous silicon flat panel imaging devices for dosimetry purposesMedical Physics, 2004
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging deviceMedical Physics, 2001
- Radiation field edge detection in portal imagesPhysics in Medicine & Biology, 1991