C-arm orbits for metal artifact avoidance (MAA) in cone-beam CT
- 18 August 2020
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 65 (16), 165012
- https://doi.org/10.1088/1361-6560/ab9454
Abstract
Metal artifacts present a challenge to cone-beam CT (CBCT) image-guided surgery, obscuring visualization of metal instruments and adjacent anatomy-often in the very region of interest pertinent to the imaging/surgical tasks. We present a method to reduce the influence of metal artifacts by prospectively defining an image acquisition protocol-viz., the C-arm source-detector orbit-that mitigates metal-induced biases in the projection data. The metal artifact avoidance (MAA) method is compatible with simple mobile C-arms, does not require exact prior information on the patient or metal implants, and is consistent with 3D filtered backprojection (FBP), more advanced (e.g. polyenergetic) model-based image reconstruction (MBIR), and metal artifact reduction (MAR) post-processing methods. The MAA method consists of: (i) coarse localization of metal objects in the field-of-view (FOV) via two or more low-dose scout projection views and segmentation (e.g. a simple U-Net) in coarse backprojection; (ii) model-based prediction of metal-induced x-ray spectral shift for all source-detector vertices accessible by the imaging system (e.g. gantry rotation and tilt angles); and (iii) identification of a circular or non-circular orbit that reduces the variation in spectral shift. The method was developed, tested, and evaluated in a series of studies presenting increasing levels of complexity and realism, including digital simulations, phantom experiment, and cadaver experiment in the context of image-guided spine surgery (pedicle screw implants). The MAA method accurately predicted tilted circular and non-circular orbits that reduced the magnitude of metal artifacts in CBCT reconstructions. Realistic distributions of metal instrumentation were successfully localized (0.71 median Dice coefficient) from 2-6 low-dose scout views even in complex anatomical scenes. The MAA-predicted tilted circular orbits reduced root-mean-square error (RMSE) in 3D image reconstructions by 46%-70% and 'blooming' artifacts (apparent width of the screw shaft) by 20-45%. Non-circular orbits defined by MAA achieved a further similar to 46% reduction in RMSE compared to the best (tilted) circular orbit. The MAA method presents a practical means to predict C-arm orbits that minimize spectral bias from metal instrumentation. Resulting orbits-either simple tilted circular orbits or more complex non-circular orbits that can be executed with a motorized multi-axis C-arm-exhibited substantial reduction of metal artifacts in raw CBCT reconstructions by virtue of higher fidelity projection data, which are in turn compatible with subsequent MAR post-processing and/or polyenergetic MBIR to further reduce artifacts.Funding Information
- Siemens Healthineers
- NIH (R01-EB-017226)
This publication has 52 references indexed in Scilit:
- Automated implant segmentation in cone-beam CT using edge detection and particle countingInternational Journal of Computer Assisted Radiology and Surgery, 2013
- Normalized metal artifact reduction (NMAR) in computed tomographyMedical Physics, 2010
- Percutaneous Needle Biopsy for Small Lung Nodules Beneath the Rib Under CT Scan Fluoroscopic Guidance With Gantry TiltSocial psychiatry. Sozialpsychiatrie. Psychiatrie sociale, 2004
- Using a flat-panel detector in high resolution cone beam CT for dental imagingDento maxillo facial radiology, 2004
- An improved exact filtered backprojection algorithm for spiral computed tomographyAdvances in Applied Mathematics, 2004
- Segmentation-free statistical image reconstruction for polyenergetic x-ray computed tomography with experimental validationPhysics in Medicine & Biology, 2003
- A general scheme for constructing inversion algorithms for cone beam CTInternational Journal of Mathematics and Mathematical Sciences, 2003
- Iterative deblurring for CT metal artifact reductionIEEE Transactions on Medical Imaging, 1996
- Penalized weighted least-squares image reconstruction for positron emission tomographyIEEE Transactions on Medical Imaging, 1994
- A Method for Correcting Bone Induced Artifacts in Computed Tomography ScannersJournal of Computer Assisted Tomography, 1978