Corrosion of titanium implants and connected prosthetic alloys using lactic acid immersion test

Abstract
Background/purpose: Combination of dental titanium implants with other prosthetic metallic components may lead to metal ion release that increases the risk of adverse reactions in patients. The present study therefore aimed to determine in vitro metal ion dissolution from different alloy combinations. Materials and methods: Dental alloys were subjected to a lactic acid immersion test together with titanium implants and matched with controls. Between day 1 and 38, open direct current potentials (DCP) between the samples and the electrolyte were recorded and metal dissociation inside the electrolyte was assessed using ICP-MS. Results: Absolute DCPs of the different alloys increased significantly (p