Abstract
Background Multiple left ventricular pacing strategies have been suggested for improving response to cardiac resynchronization therapy (CRT). However, these programming strategies may sometimes entail accepting configurations with high pacing threshold and accelerated battery drain. We assessed the feasibility of predefined pacing programming protocols, and we evaluated their impact on device longevity and their cost-impact. Methods We estimated battery longevity in 167 CRT-D patients based on measured pacing parameters according to multiple alternative programming strategies: single-site pacing associated with lowest threshold, non-apical location, longest interventricular delay, and pacing from two electrodes. To determine the economic impact of each programming strategy, we applied the results of a model-based cost analysis using a 15-year time horizon. Results Selecting the electrode with the lowest threshold resulted in a median device longevity of 11.5 years. Non-apical pacing and interventricular delay maximization were feasible in most patients and were obtained at the price of a few months of battery life. Device longevity of > 10 years was preserved in 87% of cases of non-apical pacing and in 77% on pacing at the longest interventricular delay. The mean reduction in battery life when the second electrode was activated was 1.5 years. Single-site pacing strategies increased the therapy cost by 4–6%, and multi-site pacing by 12–13%, in comparison with the lowest-cost scenario. Conclusions Modern CRT-D systems ensure effective pacing and allow multiple optimization strategies for maximizing service life or for enhancing effectiveness. Single- or multi-site pacing strategies can be implemented without compromising device service life and at an acceptable increase in therapy cost.Keywords
This publication has 32 references indexed in Scilit:
- Impact of extending device longevity on the long-term costs of implantable cardioverter-defibrillator therapy: a modelling study with a 15-year time horizonEP Europace, 2013
- Acute haemodynamic comparison of multisite and biventricular pacing with a quadripolar left ventricular leadEP Europace, 2013
- Clinical and procedural outcome of patients implanted with a quadripolar left ventricular lead: Early results of a prospective multicenter studyHeart Rhythm, 2012
- Sites of left and right ventricular lead implantation and response to cardiac resynchronization therapy observations from the REVERSE trialEuropean Heart Journal, 2012
- The relationship between ventricular electrical delay and left ventricular remodelling with cardiac resynchronization therapyEuropean Heart Journal, 2011
- Management of Phrenic Stimulation in CRT Patients over the Long Term: Still an Unmet Need ?Pacing and Clinical Electrophysiology, 2011
- Left Ventricular Lead Position and Clinical Outcome in the Multicenter Automatic Defibrillator Implantation Trial–Cardiac Resynchronization Therapy (MADIT-CRT) TrialCirculation, 2011
- Left ventricular pacing with a new quadripolar transvenous lead for CRT: Early results of a prospective comparison with conventional implant outcomesHeart Rhythm, 2011
- Insights From a Cardiac Resynchronization Optimization Clinic as Part of a Heart Failure Disease Management ProgramJournal of the American College of Cardiology, 2009
- The Growing Mismatch Between Patient Longevity and the Service Life of Implantable Cardioverter-DefibrillatorsJournal of the American College of Cardiology, 2005