Development and validation of a novel fibre-optic respiratory rate sensor (FiRRS) integrated in oxygen delivery devices
Open Access
- 3 December 2020
- journal article
- research article
- Published by IOP Publishing in Journal of Physics D: Applied Physics
- Vol. 54 (12), 124002
- https://doi.org/10.1088/1361-6463/abd061
Abstract
Respiratory rate provides crucial information on the overall health condition of patients and a reliable, low cost respiratory rate monitor for normal hospital inpatient or home use would be of significant benefit. The proposed system measures light reflection from a Fibre Bragg Grating located near, and the total reflection spectrum from a humidity sensing film deposited at, the tip of an optical fibre. Every breath causes a shift in the wavelength reflected from the Fibre Bragg Grating and intensity change in the overall reflection spectrum. The accuracy of different techniques is investigated in a two-part study with fifteen healthy volunteers. In Part 1, the participants' respiration rate followed a handheld mobile application at 5, 12 and 30 breaths per minute with simultaneous measurement using the optical fibre system, thoracic impedance pneumography and capnometry device (where possible). Two types of medical face masks and a nasal cannula with oxygen delivery rates were investigated. In Part 2, participants wore an anaesthetic face mask and breathed at normal and low tidal volumes to evaluate whether low tidal volumes could be detected. The most accurate measurement of respiratory rate was through monitoring the Bragg wavelength shift (mean accuracy = 88.1%), followed by the intensity change at the Bragg wavelength (mean accuracy = 78.9%), capnometry (mean accuracy = 77.8%), area under the overall spectrum (mean accuracy = 65.4%) and thoracic impedance pneumography (mean accuracy = 43.1%). The Fibre-optic Respiratory Rate Sensor system can differentiate between normal and low tidal volumes (p-value<0.05) and demonstrated higher accuracy than capnometry measurement of end-tidal carbon dioxide in exhaled air. These latter two monitors measured respiratory rate more accurately than thoracic impedance pneumography. A comparable accuracy in the measurement of respiratory rate was obtained when the Fibre-optic Respiratory Rate Sensor system was implemented in nasal cannula and face masks.Funding Information
- Engineering and Physical Sciences Research Council (EP/N026985/1)
This publication has 32 references indexed in Scilit:
- Association of overweight and obesity with obstructive sleep apnoea: A systematic review and meta-analysisObesity Medicine, 2020
- Recent development of respiratory rate measurement technologiesPhysiological Measurement, 2019
- Respiratory rate detection by a time-based measurement systemRevista Mexicana de Ingeniería Biomédica, 2016
- An assessment of algorithms to estimate respiratory rate from the electrocardiogram and photoplethysmogramPhysiological Measurement, 2016
- Relation between respiratory rate and heart rate – A comparative studyIndian Journal of Clinical Anatomy and Physiology, 2016
- Improvement of Dynamic Respiration Monitoring Through Sensor Fusion of Accelerometer and Gyro-sensorJournal of Electrical Engineering & Technology, 2014
- Photoplethysmographic derivation of respiratory rate: a review of relevant physiologyJournal of Medical Engineering & Technology, 2011
- Obstructive sleep apnoea and anaesthesiaAnaesthesia & Intensive Care Medicine, 2011
- Epidemiology of Obstructive Sleep ApneaAmerican Journal of Respiratory and Critical Care Medicine, 2002
- Multifactorial Risk Index for Predicting Postoperative Respiratory Failure in Men After Major Noncardiac SurgeryAnnals of Surgery, 2000