A New Test Method for In Vitro Evaluation of Pedicle Screw Loosening Potential
- 18 October 2021
- journal article
- research article
- Published by ASME International in Journal of Medical Devices
- Vol. 15 (4)
- https://doi.org/10.1115/1.4052517
Abstract
This paper proposes a new testing method based on the toggle effect under transverse loads (cranial-caudal) to investigate the loosening potential of pedicular screw designs. A three-step in vitro testing procedure was developed to mimic the loosening mechanism of pedicular screws. Firstly, the pedicular screw of a certain design is inserted into a bone substitute model specifically designed for the test. Secondly, a controlled cyclic cranial-caudal loading is applied transversally to the longitudinal axis of the screw for three ascendent load levels (staircase) by a pre-determined number of load cycles. Lastly, each pedicular screw is adjusted and submitted to axial pull-out quasi-static testing. The results are used to calculate a loosening index that, together with statistical analysis, indicates the potential for loosening of the specific design evaluated. The proposed testing method effectively provides a simulated environment to evaluate the loosening potential of pedicular screw designs. The proposed loosening index calculation may be used to compare different pedicular screw designs. The proposed methodology was verified as a valuable tool to investigate the influence of the cranial-caudal loads on pedicular screw behavior. It offers a new alternative for use in pre-clinical studies on the loosening potential of pedicular screw designs.Keywords
This publication has 23 references indexed in Scilit:
- Pedicle screw loosening: a clinically relevant complication?European Spine Journal, 2015
- Effect of Physiological Loads on Cortical and Traditional Pedicle Screw FixationSpine, 2014
- Lumbar Transpedicular Implant Failure: A Clinical and Surgical Challenge and Its Radiological AssessmentAsian Spine Journal, 2014
- The Effect of “Toggling” on the Pullout Strength of Bone Screws in Normal and Osteoporotic Bone ModelsThe Open Mechanical Engineering Journal, 2013
- Validity of synthetic bone as a substitute for osteoporotic cadaveric femoral heads in mechanical testingBone & Joint Research, 2012
- Pedicle screw loosening in dynamic stabilization: incidence, risk, and outcome in 126 patientsNeurosurgical Focus, 2011
- Screw loosening in the Dynesys stabilization system: radiographic evidence and effect on outcomesNeurosurgical Focus, 2010
- Biomechanical Analysis of Different Techniques in Revision Spinal InstrumentationSpine, 2008
- 1997 Volvo Award Winner in Basic Science StudiesSpine, 1997
- Loads on an internal spinal fixation device during walkingJournal of Biomechanics, 1997