Background Initial stability of uncemented acetabular components in total hip arthroplasty (THA) is important for osseointegration and potentially enhanced by screw fixation. We used Australian Orthopaedic Association National Joint Replacement Registry data to determine whether screw usage influences uncemented acetabular component survival. Methods Primary THA with uncemented acetabular components performed for osteoarthritis from 1999 to 2018 was included. Survivorship was calculated using Kaplan-Meier estimates of cumulative percent revision (CPR). Comparisons used Cox proportional hazards method. An instrumental variable analysis adjusted for surgeon preference for screws as a confounding factor was used. Results Three hundred thirty thousand one hundred ninety-two THAs were included (31.8% with screws, 68.2% without). Two hundred twenty thousand six hundred seven were included in the instrumental variable analysis. Revision rate of acetabular components (all causes) was higher with screws during the first six years (hazard ratio (HR) = 1.45 (95% CI 1.34, 1.57), p < 0.001) and lower thereafter (HR = 0.81 (95% CI 0.67, 0.98), p = 0.027). Revision rate of acetabular components for loosening was higher with screws over the entire study period (HR = 1.73 (95% CI 1.51, 1.98), p < 0.001). Overall THA revision rate was higher with screws during the first six years (HR = 1.20 (95% CI 1.15, 1.26), p < 0.001) but lower thereafter (HR = 0.89 (95% CI 0.81, 0.98), p = 0.020). Revision rate for dislocation was higher with screws over the entire period (HR = 1.16 (95% CI 1.06, 1.26), p < 0.001). Instrumental variable analysis revealed higher revision rates with acetabular screws in the first six years. (HR = 1.18 (95% CI 1.09–1.29), p < 0.001). Conclusion Screws did not confer a protective effect against acetabular loosening and were not associated with long-term negative consequences.Keywords
This publication has 24 references indexed in Scilit:
- Effect of screw fixation on acetabular component alignment change in total hip arthroplastyInternational Orthopaedics, 2014
- Initial Stability of Press-Fit Acetabular Components Under Rotational ForcesThe Journal of Arthroplasty, 2013
- Press-fit cementless acetabular fixation with and without screwsInternational Orthopaedics, 2013
- A cementless, elastic press-fit socket with and without screwsActa Orthopaedica, 2012
- The number of screws, bone quality, and friction coefficient affect acetabular cup stabilityMedical Engineering & Physics, 2007
- Porous-coated cups fixed with screws: A 12-year clinical and radiostereometric follow-up study of 50 hipsActa Orthopaedica, 2006
- Effect of augmented cup fixation on stability, wear, and osteolysis: A 5-year follow-up of total hip arthroplasty with RSAThe Journal of Arthroplasty, 2004
- Hydroxyapatite and tricalcium phosphate–coated cups with and without screw fixation: A randomized study of 64 hipsThe Journal of Arthroplasty, 2000
- A quantitative study of bone and soft tissues in cementless porous-coated acetabular components retrieved at autopsyThe Journal of Arthroplasty, 1993
- Observations on the Effect of Movement on Bone Ingrowth into Porous-Surfaced ImplantsClinical Orthopaedics and Related Research, 1986