EVALUATING THE DYNAMIC PERFORMANCE OF INTERFACIAL PRESSURE SENSORS AT A SIMULATED BODY-DEVICE INTERFACE
Open Access
- 4 January 2021
- journal article
- Published by University of Toronto Libraries - UOTL in CANADIAN PROSTHETICS & ORTHOTICS JOURNAL
- Vol. 4 (1)
- https://doi.org/10.33137/cpoj.v4i1.36059
Abstract
BACKGROUND: Pressure sensing at the body-device interface can help assess the quality of fit and function of assistive devices during physical activities and movement such as walking and running. However, the dynamic performance of various pressure sensor configurations is not well established. OBJECTIVE(S): Two common commercially available thin-film pressure sensors were tested to determine the effects of clinically relevant setup configurations focusing on loading areas, interfacing elements (i.e. ‘puck’) and calibration methods. METHODOLOGY: Testing was performed using a customized universal testing machine to simulate dynamic, mobility relevant loads at the body-device interface. Sensor performance was evaluated by analyzing accuracy and hysteresis. FINDINGS: The results suggest that sensor calibration method has a significant effect on sensor performance although the difference is mitigated by using an elastomeric loading puck. Both sensors exhibited similar performance during dynamic testing that agree with accuracy and hysteresis values reported by manufacturers and in previous studies assessing mainly static and quasi-static conditions. CONCLUSION: These findings suggest that sensor performance under mobility relevant conditions may be adequately represented via static and quasi-testing testing. This is important since static testing is much easier to apply and reduces the burden on users to verify dynamic performance of sensors prior to clinical application. The authors also recommend using a load puck for dynamic testing conditions to achieve optimal performance. Layman's Abstract Pressure sensors can be used in prosthetics to provide clinicians with data about how well a device fits and functions. However, pressure sensors are unproven when it comes to use during activities such as walking or running. This study tested two common pressure sensors in a setup that applied forces similar to walking. These findings indicate that sensor calibration affects sensor accuracy. Accuracy can be improved by applying a small puck to the sensor to spread the load more evenly. With the puck, the performance of the sensors was found to be acceptable for potential use in clinical applications. These findings also show that dynamic testing of pressure sensors may not be needed prior to clinical usage. Instead, performance can be based on static testing which is easier to do. Article PDF Link: https://jps.library.utoronto.ca/index.php/cpoj/article/view/36059/27891 How To Cite: Hamilton M, Sivasambu H, Behdinan K, Andrysek J. Evaluating the dynamic performance of interfacial pressure sensors at a simulated body-device interface. Canadian Prosthetics & Orthotics Journal. 2021;Volume 4, Issue 1, No.4. https://doi.org/10.33137/cpoj.v4i1.36059 Corresponding Author: Jan Andrysek, PhD, Bloorview Research Institute, Holland Bloorview Kids Rehabilitation Hospital, Toronto, Canada. Email: jandrysek@hollandbloorview.ca ORCID: https://orcid.org/0000-0002-4976-1228Keywords
This publication has 11 references indexed in Scilit:
- Flexible Sensors for Pressure Therapy: Effect of Substrate Curvature and Stiffness on Sensor PerformanceSensors, 2017
- Evaluating and improving the performance of thin film force sensors within body and device interfacesMedical Engineering & Physics, 2017
- Evaluation of Flexible Force Sensors for Pressure Monitoring in Treatment of Chronic Venous DisordersSensors, 2017
- The effect of biomechanical variables on force sensitive resistor error: Implications for calibration and improved accuracyJournal of Biomechanics, 2016
- Gait Analysis Methods: An Overview of Wearable and Non-Wearable Systems, Highlighting Clinical ApplicationsSensors, 2014
- Dynamic interface pressure distributions of two transtibial prosthetic socket concepts.2009
- Predictive modelling of human walking over a complete gait cycleJournal of Biomechanics, 2007
- Anatomical hip model for the mechanical testing of hip protectorsMedical Engineering & Physics, 2005
- Scientific validation of two commercial pressure sensor systems for prosthetic socket fitProsthetics and Orthotics International, 2000
- A new biomechanical method for determination of static prosthetic alignmentProsthetics and Orthotics International, 1997