Evaluation of a multiple breath nitrogen washout system in children
- 21 May 2020
- journal article
- research article
- Published by Wiley in Pediatric Pulmonology
- Vol. 55 (8), 2108-2114
- https://doi.org/10.1002/ppul.24862
Abstract
Introduction The multiple breath nitrogen washout (MBW) test offers a sensitive measure of airway function. In this study we aim to (a) assess the validity of the EasyOne Pro LAB (MBWndd) in an in vitro lung model, (b) assess the feasibility, repeatability, and reproducibility of MBWndd and (c) compare outcomes with the Exhalyzer D (MBWEM) and body plethysmography. Methods In vitro, functional residual capacity (FRC) measurements were assessed using a lung model under quasi-physiological conditions and compared to measured FRC. In vivo plethysmography and MBW were performed in a prospective study of children at two visits (n = 45 healthy; n = 41 cystic fibrosis [CF]). Bland-Altman plots were used to compare agreement between FRC and lung clearance index (LCI) measurements. Results In vitro FRCndd measurements were repeatable but lung volumes were underestimated (mean relative difference −5.4% (limits of agreement [LA] −9.6%; −1.1%), 95% confidence interval (CI) −6.27; −4.45). In vivo, compared to plethysmography, FRCndd was consistently lower (−19.3% [−40.5; 1.9], 95% CI [−23.9; −14.7]), and showed a volume dependency. LCIndd values were also higher in children with smaller lung volumes. The within-test coefficient of variation of the FRCndd and LCIndd were 4.9% in health, and 5.6% and 6.9% in CF respectively. LCIndd was reproducible between-visits (mean relative difference [LA] −3.7% [−14.8, −7.5; 95% CI −6.6; −0.73] in health [n = 17] and 0.34% [−13.2, 22.8; 95% CI −5.0; 5.69] in CF [n = 23]). When calculated using the same algorithm, LCIndd was similar to LCIEM in health. Conclusions MBWndd measurements are feasible, repeatable, and reproducible, however, MBW-derived outcomes are not interchangeable with MBWEM.Keywords
Funding Information
- Vertex Pharmaceuticals
This publication has 18 references indexed in Scilit:
- Validation of multiple‐breath washout equipment for infants and young childrenPediatric Pulmonology, 2014
- Validation of a Photoacoustic Gas Analyser for the Measurement of Functional Residual Capacity Using Multiple-Breath Inert Gas WashoutRespiration, 2014
- Multiple Breath Nitrogen Washout: A Feasible Alternative to Mass SpectrometryPLOS ONE, 2013
- Consensus statement for inert gas washout measurement using multiple- and single- breath testsEuropean Respiratory Journal, 2013
- Multi-ethnic reference values for spirometry for the 3–95-yr age range: the global lung function 2012 equationsEuropean Respiratory Journal, 2012
- A Realistic Validation Study of a New Nitrogen Multiple-Breath Washout SystemPLOS ONE, 2012
- Feasibility and Variability of Measuring the Lung Clearance Index in a Multi‐Center SettingPediatric Pulmonology, 2011
- Lung clearance index: Normal values, repeatability, and reproducibility in healthy children and adolescentsPediatric Pulmonology, 2009
- Standardisation of the measurement of lung volumesEuropean Respiratory Journal, 2005
- Statistics notes: Measurement errorBMJ, 1996