Bacterial contamination of ultrasound probes and coupling gels in a university hospital in Turkey
Open Access
- 4 July 2019
- journal article
- Published by The European Research Journal in The European Research Journal
- Vol. 5 (4), 582-587
- https://doi.org/10.18621/eurj.401327
Abstract
Objectives: Nosocomial outbreaks of infection originating from ultrasound probes and contaminated coupling gels have been reported. It was reported that the ultrasound probe, if cultured after routine scanning of intact skin, may become colonized with skin flora. Methods: Culture swabs from 22 probes of the 9 ultrasound machines and from the gels in the 10 gel folders were taken. All swabs taken from probe head, probe holder and the coupling gel in the folder at the beginning of the day were cultured. After fifth scanning and after wiping off the gel with a dry, nonsterile paper towel, cultures were again obtained from probe head and probe holder. Results: A total of 98 culture results were included of which 42.8% were positive for bacterial growth. The rate of bacterial contamination from probes at morning before the start of examination and after scanning were 34.1% and 56.8%, respectively and this difference was statistically significant (p = 0.023). Conclusions: We think that using nonsterile, dry, soft and absorbent paper towel after each procedure, could be inadequate for disinfection of probe head. Especially, good hand hygiene could decrease the rate of growth of bacterial colony at probe handle.Keywords
This publication has 19 references indexed in Scilit:
- Evaluation of a new disinfection procedure for ultrasound probes using ultraviolet lightJournal of Hospital Infection, 2007
- An investigation of the microbiological contamination of ultrasound equipmentBritish Journal of Infection Control, 2006
- Burkholderia cepaciaInfections Associated With Intrinsically Contaminated Ultrasound Gel: The Role of Microbial Degradation of ParabensInfection Control & Hospital Epidemiology, 2004
- Infection Control — A Problem for Patient SafetyThe New England Journal of Medicine, 2003
- Bacterial Colonisation of Doppler Probes on Vascular Surgical WardsEuropean Journal of Vascular and Endovascular Surgery, 2002
- An Outbreak of Pyodermas Among Neonates Caused by Ultrasound Gel Contaminated With Methicillin-SusceptibleStaphylococcus AureusInfection Control & Hospital Epidemiology, 2000
- Letters to the EditorJournal of Hospital Infection, 1999
- Microbiologic Assessment of the Transabdominal Ultrasound Transducer HeadSouthern Medical Journal, 1996
- Can ultrasound probes and coupling gel be a source of nosocomial infection in patients undergoing sonography? An in vivo and in vitro study.American Journal of Roentgenology, 1995
- Ultrasound scanning of post-operative wounds — the risks of cross-infectionClinical Radiology, 1988