Investigation of potential safety hazards during medical waste disposal in SARS-CoV-2 testing laboratory
Open Access
- 6 March 2021
- journal article
- research article
- Published by Springer Science and Business Media LLC in Environmental Science and Pollution Research
- Vol. 28 (27), 35822-35829
- https://doi.org/10.1007/s11356-021-13247-4
Abstract
This study aims to investigate the potential safety hazards and provide reference for improving the medical waste disposal procedure in SARS-CoV-2 testing laboratory. Our SARS-CoV-2 testing group detected the RNA residue on the surface of medical waste with Droplet Digital PCR, and held a meeting to discuss the risks in the laboratory medical waste disposal process. After effective autoclaving, SARS-CoV-2 contaminated on the surface of medical waste bags was killed, but the average concentration of viral RNA residues was still 0.85 copies/cm2. It would not pose a health risk, but might contaminate the laboratory and affect the test results. When the sterilized medical waste bags were transferred directly by the operators without hand disinfection, re-contamination would happen, which might cause the virus to leak out of the laboratory. Furthermore, we found that sterilization effect monitoring and cooperation among operators were also very important. In summary, we investigated and analyzed the potential safety hazards during the medical waste disposal process in SARS-CoV-2 testing laboratory, and provided reasonable suggestions to ensure the safety of medical waste disposal.Funding Information
- Fundamental Research Funds for the Central Universities (2662017PY011)
This publication has 29 references indexed in Scilit:
- Effectiveness of autoclaving in sterilizing reusable medical devices in healthcare facilitiesJournal of Infection in Developing Countries, 2019
- Validation of Autoclave Protocols for Successful Decontamination of Category A Medical Waste Generated from Care of Patients with Serious Communicable DiseasesJournal of Clinical Microbiology, 2017
- A Preliminary Evaluation of a Reusable Digital Sterilization Indicator PrototypeThe Journal of Contemporary Dental Practice, 2014
- Persistence of viral genomes after autoclavingJournal of Virological Methods, 2014
- Short RNA indicator sequences are not completely degraded by autoclavingScientific Reports, 2014
- Developing the Use of Quality Indicators in Sterilization Practices2012
- UV irradiation and autoclave treatment for elimination of contaminating DNA from laboratory consumablesForensic Science International: Genetics, 2010
- Comparison of a rapid readout biological indicator for steam sterilization with four conventional biological indicators and five chemical indicators.1996
- Comparison of a Rapid Readout Biological Indicator for Steam Sterilization with Four Conventional Biological Indicators and Five Chemical IndicatorsInfection Control & Hospital Epidemiology, 1996
- COMPARATIVE THERMORESISTANCE OF TWO BIOLOGICAL INDICATORS FOR MONITORING STEAM AUTOCLAVESActa Pathologica Microbiologica Scandinavica Series B: Microbiology, 1986