Improvement of megavoltage computed tomography image quality for adaptive helical tomotherapy using cycleGAN‐based image synthesis with small datasets
- 21 August 2021
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 48 (10), 5593-5610
- https://doi.org/10.1002/mp.15182
Abstract
Purpose Megavoltage computed tomography (MVCT) offers an opportunity for adaptive helical tomotherapy. However, high noise and reduced contrast in the MVCT images due to a decrease in the imaging dose to patients limits its usability. Therefore, we propose an algorithm to improve the image quality of MVCT. Methods The proposed algorithm generates kilovoltage CT (kVCT)-like images from MVCT images using a cycle-consistency generative adversarial network (cycleGAN)-based image synthesis model. Data augmentation using an affine transformation was applied to the training data to overcome the lack of data diversity in the network training. The mean absolute error (MAE), root-mean-square error (RMSE), peak signal-to-noise ratio (PSNR), and structural similarity index measure (SSIM) were used to quantify the correction accuracy of the images generated by the proposed algorithm. The proposed method was validated by comparing the images generated with those obtained from conventional and deep learning-based image processing method through non-augmented datasets. Results The average MAE, RMSE, PSNR, and SSIM values were 18.91 HU, 69.35 HU, 32.73 dB, and 95.48 using the proposed method, respectively, whereas cycleGAN with non-augmented data showed inferior results (19.88 HU, 70.55 HU, 32.62 dB, 95.19, respectively). The voxel values of the image obtained by the proposed method also indicated similar distributions to those of the kVCT image. The dose-volume histogram of the proposed method was also similar to that of electron density corrected MVCT. Conclusions The proposed algorithm generates synthetic kVCT images from MVCT images using cycleGAN with small patient datasets. The image quality achieved by the proposed method was correspondingly improved to the level of a kVCT image while maintaining the anatomical structure of an MVCT image. The evaluation of dosimetric effectiveness of the proposed method indicates the applicability of accurate treatment planning in adaptive radiation therapy.Keywords
This publication has 56 references indexed in Scilit:
- Megavoltage CT imaging quality improvement on TomoTherapy via tensor frameletMedical Physics, 2013
- Low-dose CT reconstruction via edge-preserving total variation regularizationPhysics in Medicine & Biology, 2011
- Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic CT images from highly undersampled projection data setsMedical Physics, 2008
- Interfractional Variations in Patient Setup and Anatomic Change Assessed by Daily Computed TomographyInternational Journal of Radiation Oncology*Biology*Physics, 2007
- Deformable registration of the planning image (kVCT) and the daily images (MVCT) for adaptive radiation therapyPhysics in Medicine & Biology, 2006
- Artifacts in CT: Recognition and AvoidanceRadioGraphics, 2004
- Image Quality Assessment: From Error Visibility to Structural SimilarityIEEE Transactions on Image Processing, 2004
- Multimodality image registration by maximization of mutual informationIEEE Transactions on Medical Imaging, 1997
- Computed tomography with a linear accelerator with radiotherapy applicationsMedical Physics, 1983
- A 4-MV CT scanner for radiation therapy: The prototype systemMedical Physics, 1982