Wireless 3D Surgical Navigation and Tracking System With 100μm Accuracy Using Magnetic-Field Gradient-Based Localization

Abstract
This paper describes a high-resolution 3D navigation and tracking system using magnetic field gradients, that can replace X-Ray fluoroscopy in high-precision surgeries. Monotonically varying magnetic fields in X, Y and Z directions are created in the field-of-view (FOV) to produce magnetic field gradients, which encode each spatial point uniquely. Highly miniaturized, wireless and battery-less devices, capable of measuring their local magnetic field, are designed to sense the gradient field. One such device can be attached to an implant inside the body and another to a surgical tool, such that both can simultaneously measure and communicate the magnetic field at their respective locations to an external receiver. The relative location of the two devices on a real-time display can enable precise surgical navigation without using X-Rays. A prototype device is designed consisting of a micro-chip fabricated in 65nm CMOS technology, a 3D magnetic sensor and an inductor-coil. Planar electromagnetic coils are designed for creating the 3D magnetic field gradients in a 20x20x10cm3 of scalable FOV. Unambiguous and orientation-independent spatial encoding is achieved by: (i) using the gradient in the total field magnitude instead of only the Z-component; and (ii) using a combination of the gradient fields to correct for the non-linearity and non-monotonicity in X and Y gradients. The resultant X and Y FOV yield ≥90% utilization of their respective coil-span. The system is tested in vitro to demonstrate a localization accuracy of <100μm in 3D, the highest reported to the best of our knowledge.
Funding Information
  • National Science Foundation (1823036)
  • Rothenberg Innovation Initiative (101170)
  • Heritage Medical Research Institute