Evaluation of Nucleoprotein-Based Enzyme-Linked Immunosorbent Assay for Serodiagnosis of Acute Crimean–Congo Hemorrhagic Fever Virus Infections in a Turkish Population

Abstract
Background: Crimean–Congo hemorrhagic fever virus (CCHFV) causes a highly contagious tick-borne disease with high case-fatality rates in humans. It is circulating not only in many Asian and African countries, but also spreading to and within Europe. To cope better with future outbreaks of Crimean–Congo hemorrhagic fever (CCHF), the WHO has prioritized the need for the development and validation of CCHF diagnostics, including serological assays. In this study, we evaluated the performance of the new EUROIMMUN anti-CCHFV IgM and IgG enzyme-linked immunosorbent assays (ELISAs). Materials and Methods: Both ELISAs were compared to the Vector-Best VectoCrimean-CHF-IgM and -IgG ELISAs using the EUROIMMUN CCHFV Mosaic 2 IgM and IgG indirect immunofluorescence assays (IFA) as reference. Forty-nine acute-phase serum samples from patients with CCHFV infection confirmed by reverse transcription-polymerase chain reaction (RT-PCR) and/or anti-CCHFV IgM IFA positivity were used to determine assay sensitivity. The assessment of specificity was based on sera from 30 control patients, 30 healthy blood donors, and 29 patients with hantavirus or sandfly fever virus infections. All samples originated from Turkey. Results: Sensitivity of the EUROIMMUN ELISAs (IgM 98.0%, IgG 47.1%) exceeded that of the Vector-Best ELISAs (IgM 95.9%, IgG 35.3%). Specificity of the EUROIMMUN ELISA IgM (86.4%) was slightly higher compared with the Vector-Best ELISA IgM (84.7%), while specificity for IgG was 100% for both assays. Qualitative agreement between the EUROIMMUN and Vector-Best ELISAs was substantial for detecting anti-CCHFV IgM (84.1%, ĸ = 0.673) and IgG (94.9%, ĸ = 0.791), whereas the quantitative results indicated a very strong positive correlation (IgM: r = 0.868, IgG: r = 0.913). Conclusion: The new EUROIMMUN anti-CCHFV ELISAs are standardized and easy-to-use tools that reliably support the identification of acute CCHF cases, and thus suitable for laboratories involved in on-site outbreak support.