Abstract
In individuals with transtibial limb loss, a contributing factor to mobility-related challenges is the disruption of biological calf muscle function due to transection of the soleus and gastrocnemius. Powered prosthetic ankles can restore primary function of the mono-articular soleus muscle, which contributes to ankle plantarflexion. In effect, a powered ankle acts like an artificial soleus (AS). However, the biarticular gastrocnemius connection that simultaneously contributes to ankle plantarflexion and knee flexion torques remains missing, and there are currently no commercially available prosthetic ankles that incorporate an artificial gastrocnemius (AG). The goal of this work is to describe the design of a novel emulator capable of independently controlling artificial soleus and gastrocnemius behaviors for transtibial prosthesis users during walking. To evaluate the emulator's efficacy in controlling the artificial gastrocnemius behaviors, a case series walking study was conducted with four transtibial prosthesis users. Data from this case series showed that the emulator exhibits low resistance to the user's leg swing, low hysteresis during passive spring emulation, and accurate force tracking for a range of artificial soleus and gastrocnemius behaviors. The emulator presented in this paper is versatile and can facilitate experiments studying the effects of various artificial soleus and gastrocnemius dynamics on gait or other movement tasks. Using this system, it is possible to address existing knowledge gaps and explore a wide range of artificial soleus and gastrocnemius behaviors during gait and potentially other activities of daily living.Keywords
Funding Information
- National Science Foundation (Graduate Research Fellowship Program)
This publication has 17 references indexed in Scilit:
- Variation in the human Achilles tendon moment arm during walkingComputer Methods in Biomechanics and Biomedical Engineering, 2016
- Biomechanics of the ankleOrthopaedics and Trauma, 2016
- A Universal Ankle–Foot Prosthesis Emulator for Human Locomotion ExperimentsJournal of Biomechanical Engineering, 2014
- Bionic ankle–foot prosthesis normalizes walking gait for persons with leg amputationProceedings. Biological sciences, 2011
- Systematic Variation of Prosthetic Foot Spring Affects Center-of-Mass Mechanics and Metabolic Cost During WalkingIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2011
- Mechanical energetic contributions from individual muscles and elastic prosthetic feet during symmetric unilateral transtibial amputee walking: A theoretical studyJournal of Biomechanics, 2007
- Contributions of the individual ankle plantar flexors to support, forward progression and swing initiation during walkingJournal of Biomechanics, 2001
- Grade III open tibial fractures: Functional outcome and quality of life in amputees versus patients with successful reconstructionInjury, 2001
- Muscle balance at the knee-moment arms for the normal knee and the ACL-minus kneeIEEE Transactions on Rehabilitation Engineering, 1997
- An estimation of power output and work done by the human triceps surae musle-tendon complex in jumpingJournal of Biomechanics, 1986