Wavelength-Filter Based Spectral Calibrated Wave number - Linearization in 1.3 mm Spectral Domain Optical Coherence

  • 1 December 2013
    • Vol. 3 (2), 336-340
Abstract
In this study, we demonstrate the enhanced spectral calibration method for 1.3 μm spectral-domain optical coherence tomography (SD-OCT). The calibration method using wavelength-filter simplifies the SD-OCT system, and also the axial resolution and the entire speed of the OCT system can be dramatically improved as well. An externally connected wavelength-filter is utilized to obtain the information of the wavenumber and the pixel position. During the calibration process the wavelength-filter is placed after a broadband source by connecting through an optical circulator. The filtered spectrum with a narrow line width of 0.5 nm is detected by using a line-scan camera. The method does not require a filter or a software recalibration algorithm for imaging as it simply resamples the OCT signal from the detector array without employing rescaling or interpolation methods. One of the main drawbacks of SD-OCT is the broadened point spread functions (PSFs) with increasing imaging depth can be compensated by increasing the wavenumber-linearization order. The sensitivity of our system was measured at 99.8 dB at an imaging depth of 2.1 mm compared with the uncompensated case.